For further information:
PCS2000
Features of
PCS2000
About
PCS2000
PCS2000
Software
FPM1
Download Product
Information
Monitor Specification
Sensor Head Specification
|
|
Sensor Head Specification
- Base size ranges: 2-3,3-5,5-10,10-15,15-20,20-25,
>2µm, >25µm
- Option: 3-7µm
- Sensing method: Laser obscuration method
- Laser wavelength: 780nm, power 3mW
- Counting method: High density PAL with processor
interrogation
- Statistical basis: Homogeneous distribution assumed,
so that fractional volume measurement is valid. Baysian
update of best estimate
- Coincidence limit: 33,000 particles per ml
- Counting accuracy: +/-10% at half scale, +/-20% at
full scale
- Size accuracy: +/-10% for 5µm latex spheres
- Sample time: 20 seconds or 100 seconds
- Flow rate: Typically 20 ml/min (between 10 and 60
ml/min)
- Calcification: Automatic compensation. Error warning
if too great
- Cleaning: Simple pipe brush
- Calibration verification: 8 point, built in
- Sensor address: 4 way switch
- Power: 24V supplied from monitor
- Enclosure: Aluminium with 1/4” BSP fittings
- Size and weight: 230x140x70mm / 1.2kg
- Labelling: Laser class and wavelength warning
Monitor Specification
- Display: 20x4 character backlit LCD. LED bar graph
status for alarms
- Particle display: Total particle density per ml,
Total absolute particle count, Particle density for each
collection pot
- Alarms: Individually settable for each node
- Alarm levels: Warning and full alarm levels
- Alarm Delay: Settable 1 to 99 minutes
- Sensor detection: Automatically scannned
- Sensor control: Individual sensors can be disabled
- Log reduction: Option for dual sensor systems
- Logging database: Last 2000 samples re-circulating
- Logging: Variable 1 to 999 minutes
- Clock: 1 second resolution; year 2000 compliant
- Enclosure: Robust case with latched transparent front
cover
- Interface: Outputs: 2 x 4-20mA programmable
- Interface options: RS232 and modem options
- PC software option: Collect and display data
- PC control option: Activate all instrument functions
remotely
- Power: 85V to 265V ac or 24V dc
- Size and weight: 200x300x185mm
- Environmental: IP65, 0-30C operating, 90% RH
non-condensing
- Warranty: 12 months
- Approvals: CE
Ordering Options
- Data monitor: PCS-M
- Data monitor with modem: PCS-M-MOD
- Data monitor with log reduction: PCS-M-L
- For dc option add “24V” to any of the above e.g.
PCS-M-24V
- Slow response sensor head: PCS-S
- Slow response head with pressure sensor input:
PCS-S-P
- High speed sensor head: PCS-H
- High speed head with pressure sensor input: PCS-H-P
- Serial software: PCS-PC-Serial
- Remote software: PCS-PC-Remote
DIVERSE supplies a variety of instruments and consultancy